Description
Increase Production Efficiency by Integrity Testing Multiple Filter Assemblies Simultaneously
The Palltronic Compact Touch MUX extension enables a single Palltronic Compact Touch integrity test device to perform a Pall Pressure Decay Test on up to four filter assemblies simultaneously.
The test unit is a small, lightweight, portable instrument weighing 7 kg (15.4 lbs.). Data input to the MUX Extension instrument is done via the Compact Touch integrity test device touchscreen with all essential information displayed. The integrity test unit stabilizes the test pressure and measures pressure decay for a defined time.
The test device is dust-proof and splash-proof in accordance with IP55 rating providing protection against severe laboratory and production conditions. The venting system includes external vent valves to prevent fluid contamination backing into the unit on completion of the filter integrity test.
The MUX extension reduces pressure hold test time of up to 4 identical filter assemblies by ≥50 %. By utilizing two units in combination with one Compact Touch up to 8 identical filter assemblies may be tested.
CE Declaration
The Palltronic Compact Touch MUX Integrity Test Instrument Extension is compliant to:
- European Directive 2014/30/EU (formerly 2004/108/EC) – Electromagnetic Compatibility
- European Directive 2014/95/EU (formerly 2006/95/EC) – Low Voltage Safety
- European Directive 2011/65/EU – Restriction of Hazardous Substances (RoHS2)
- European Directive 2012/19/EU – Waste Electrical and Electronic Equipment (WEEE)
Specifications
Physical Dimensions | 250 (w) x 280 (h) x 220 (d) mm (9.8” x 11” x 8.7”) |
Weight | Unit without casing 7 kg (15.4 lbs.) |
Filter Tests | Pressure Decay Test |
Operating System | Linux* |
Test Program Storage | Up to 100 programs |
Test Result Storage | Up to 1,000 tests |
Languages Display and Input | English, German, French, Spanish, Italian, Portuguese, Brazilian |
Languages Display Only | Chinese, Japanese, Russian |
Communication Ports | CAN bus for product extensions |
Pressure Measurement Measuring Range | 50 - 6000 mbar (0.73 – 87 psi) |
Pressure Measurement Accuracy | ± 0.5 % of full scale (6000 mbar) |
Resolution | 1 mbar (0.0145 psi) |
Calibration Limits | ± 0.33 % of full scale (6000 mbar) |
Electrical Data | Voltage automatically adjusted between 100 – 240 V AC Input Frequency 50 Hz / 60 Hz External Vent Valve 24 V DC |
Monochrome Screen | Size: diagonal 6.9 cm (2.7 in) 128 x 64 pixels Illuminated Background |
Pneumatic Specifications | Maximum Gas Supply Pressure: 8,000 mbar (116 psi) Minimum Gas Supply Pressure: 1,000 mbar (14.5 psi) above test pressure Test Pressure Range: 50 – 6,500 mbar (0.7 – 94.2 psi) |
Pneumatic Connections | Compressed Air Inlet: Male Stäubli* Nipple RBE03.6150 Compressed Air Outlet: Female Stäubli Coupling RBE03.2904 Vent: Hose connection 8 mm outer diameter |
Operation and Storage Conditions | Operating Temperature: 3 °C to +50 °C (+37 °F to +122 °F) Storage Temperature: -20 °C to +70 °C (-4 °F to +158 °F) Humidity: < 95% rh (no condensation) |
IP Rating | IP55 |
ATEX | This instrument is not suitable for use in ATEX regulated atmospheres. |
Recycling Information | Please visit the Pall website for information on recycling: https://www.pall.com/en/ about-pall/corporate-sustainability/weeecompliance. html |
Type
Ordering Information
Part Number | Description |
CT001-MUX |
|
Reviews
Earn 10% off* your next order online by leaving a review of this product. Please login to your account to leave a review. We appreciate and value your feedback.
*Subject to Terms and Conditions.